FDA Brief

Company record

KENT IMAGING, INC.

0 FDA warning letters since 2023 and 1 recall in the last 12 months.

Get an email the day FDA posts anything new about KENT IMAGING, INC..

Recalls

ReportedClassFirmProduct and reason
Jul 29 Class II KENT IMAGING, INC.
Device
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.